236 VT1422A Command Reference  
ABORt
ABORt 
The ABORt subsystem is a part of the VT1422A's trigger system. ABORt resets the 
trigger system from its Wait For Trigger state to its Trigger Idle state. 
Subsystem Syntax
ABORt 
 
CAUTION
ABORT stops execution of a running algorithm. The control output is 
left at the last value set by the algorithm. Depending on the process, 
this uncontrolled situation could be dangerous. Make certain that the 
process is put into a safe state before execution of a controlling 
algorithm is halted.
Comments
ABORt does not affect any other settings of the trigger system. When the 
INITiate command is sent, the trigger system will respond just as it did before 
the ABORt command was sent.
Related Commands: INITiate[:IMMediate], TRIGger
*RST Condition: TRIG:SOUR HOLD
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
ABORT
If INITed, goes to Trigger Idle state. 
If scanning and/or running algorithms, 