VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
248 VT1422A Command Reference  
ALGorithm
When Accepted: Both before and after INIT. Also accepted before and after 
the algorithm referenced is defined.
Related Commands: ALG:UPDATE, ALG:STATE?, ALG:DEFINE
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
ALG:STATE 'ALG2',OFF
disable ALG2
ALGorithm[:EXPLicit][:STATe]?
ALGorithm[:EXPLicit][:STATe]?’<alg_name>’ 
returns the state (enabled or 
disabled) of algorithm <alg_name>.
Parameters
Comments
Since ALG:STATE is valid for an undefined algorithm, ALG:STATE? will 
return the current state for <alg_name> even if it is not currently defined.
Returned Value: Numeric, 0 or 1. Type is uint16.
*RST Condition: ALG:STATE 1
Send with VXIplug&play Function: hpe1422_cmdInt16_Q(...)
ALGorithm[:EXPLicit]:TIME?
ALGorithm[:EXPLicit]:TIME? ’<alg_name>’ 
computes and returns a worst-case 
execution time estimate in seconds.
Parameters
Comments
When <alg_name> is ALG1 through ALG32, ALG:TIME? returns only the time 
required to execute the algorithm's code.
When <alg_name> is 'MAIN', ALG:TIME? returns the worst-case execution 
time for an entire measurement & control cycle (sum of MAIN, all enabled 
algorithms, analog and digital inputs and control outputs).
If triggered more rapidly than the value returned by ALG:TIME? 'MAIN', 
the VT1422A will generate a "Trigger too fast" error.
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
alg_name
string
ALG1 - ALG32
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
alg_name
string
ALG1 - ALG32 or MAIN