268 VT1422A Command Reference  
CALibration
CALibration:SETup
CALibration:SETup 
causes the Channel Calibration function to be performed for 
every module channel with an analog SCP installed (input or output). The Channel 
Calibration function calibrates the A/D Offset and the Gain/Offset for these analog 
channels. This calibration is accomplished using internal calibration references. For 
more information see *CAL? on page 399.
Note that the scope of the *CAL? and CAL:SETup commands is limited to the 
VT1422A and the SCPs it contains. They do not calibrate Remote Signal 
Conditioning Units like the VT1529A/B. CAL:REMote? must be used in addition to 
*CAL?/CAL:SETup for RSCs.
Comments
CAL:SET performs the same operation as the *CAL? command except that, 
since it is not a query command, it doesn't tie-up the driver waiting for response 
data from the instrument. If there are multiple VT1422As in a system, a 
CAL:SET operation can be started on each and then a CAL:SET? command can 
be executed to complete the operation on each instrument.
Related Commands: CAL:SETup?, *CAL?
When Accepted: Not while INITiated
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
CAL:SET
start SCP Calibration on 1st VT1422A
:
start SCP Calibration on more VT1422As
CAL:SET
start SCP Calibration on last VT1422A
CAL:SET?
query for results from 1st VT1422A
:
query for results from more VT1422As
CAL:SET?
query for results from last VT1422A
CALibration:SETup?
CALibration:SETup? 
Returns a value to indicate the success of the last CAL:SETup 
or *CAL? operation. CAL:SETup? returns the value only after the CAL:SETup 
operation is complete.
Comments
Returned Value: 
The data type for this returned value is int16.
Value
Meaning
Further Action
0
Cal OK
None
-1
Cal Error
Query the Error Queue (SYST:ERR?)
Also run *TST?
-2
No results available
No *CAL? or CAL:SETUP done