VT1422A Command Reference   271
CALibration
Parameters
Comments
CAL:TARE also performs the equivalent of a *CAL? operation. This operation 
uses the tare constants to set a DAC which will remove each channel offset as 
"seen" by the module's A/D converter. As an example assume that the system 
wiring to channel 0 generates a +0.1 volt offset with 0 volts (a short) applied at 
the Unit Under Test (UUT). Before CAL:TARE the module would return a 
reading of 0.1 volts for channel 0. After CAL:TARE (@100), the module will 
return a reading of 0 volts with a short applied at the UUT and the system wiring 
offset will be removed from all measurements of the signal to channel 0.
The CAL:TARE command may be issued to several VT1422As to be later 
completed with a matching CAL:TARE? query sent to each instrument. Note if 
the DIAG:CAL:TARE:OTD:MODE is set to "1" then the CAL:TARE command 
will not return until the calibration is complete.
Set Amplifier/Filter SCP gain before CAL:TARE. For best accuracy, choose the 
gain that will be used during measurements. If the range or gain setup is later 
changed, be sure to perform another *CAL?.
If Open Transducer Detect (OTD) is enabled when CAL:TARE is executed, 
the module will disable OTD, wait 1 minute to allow channels to settle, perform 
the calibration and then re-enable OTD. If OTD is turned off by the program 
before executing CAL:TARE, the application should also wait one minute for 
settling. If the DIAG:CAL:TARE:OTD:MODE is set to "1", the OTD will 
remain enabled throughout the TARE calibration. This allows the voltage 
generated by the OTD current to also be removed by the TARE cal.
The maximum voltage that CAL:TARE can compensate for is dependent on the 
range chosen and SCP gain setting. The following table lists these values.
Channel offsets are compensated by the CAL:TARE command even when not 
stored in the Flash Memory. There is no need to use the CAL:STORE TARE 
command for channels which are re-calibrated frequently.
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
ch_list
channel list (string)
100 - 15731
none
Maximum CAL:TARE Offsets
A/D range
± V F.Scale
Offset V
Gain x1
Offset V
Gain x8
Offset V
Gain x16
Offset V
Gain x64
16
4
1
0.25
0.0625
3.2213
0.82101
0.23061
0.07581
0.03792
0.40104
0.10101
0.02721
0.00786
0.00312
0.20009
0.05007
0.01317
0.00349
0.00112
0.04970
0.01220
0.00297
0.00055
n/a