274 VT1422A Command Reference  
CALibration
<ref_ohms> must be within 5% of the 7500
Ω nominal reference resistor value 
or a -222 ’Data out of range’ error will be generated. If this error occurs, 
verify the external measurement equipment and run *TST? on the VT1422A.
<ref_ohms> may be specified in kOhm (kohm).
When Accepted: Not while INITiated
Related Commands: CAL:CONF:RES, CAL:STORE ADC
Send with VXIplug&play Function: hpe1422_cmd(...)
Command
Sequence
CAL:CONF:RES
(now measure ref resistor with external DMM)
CAL:VAL:RES  <measured value>
Send measured value to module
CALibration:VALue:VOLTage
CALibration:VALue:VOLTage  <ref_volts
sends the value of the just-measured dc 
reference source to the module for A/D calibration.
Parameters
Comments
The "expected" output values for the voltage reference source is:
0.9 * Nominal Range Value for the 0.0625 through 4 volt ranges.
10 volts for the 16 volt range.
Use the CAL:CONF:VOLT command to configure the on-board voltage source 
for measurement by an external reference voltmeter via the Calibration Bus 
terminals.
 A measurement of the source voltage can be made with an external multimeter 
connected to the HCAL and LCAL terminals on the Terminal Module or the 
V H and  V L terminals on the Cal Bus connector when not using a terminal 
module.
The <ref_volts> value given must be for the currently configured range and 
output (zero or full scale) as set by the previous CAL:CONF:VOLT <range>, 
ZERO | FSCale
 command.
<ref_volts> must be within 4% of the actual reference voltage value as read after 
CAL:CONF:VOLT or an error 3042 ’0x400: DSP-DAC adjustment went to 
limit’ will be generated. If the reading on the external reference voltmeter is in 
excess of 4% error from nominal voltage, verify the voltmeter and execute 
*TST? on the VT1422A.
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
ref_volts
numeric (float32)
must be within +10% or -50% of the 
"expected" source output value
volts