VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
VT1422A Command Reference   291
FETCh?
FETCh?
Subsystem Syntax
FETCh? 
returns readings stored in VME memory. 
Comments
This command is only available in systems using an Agilent/HP E1405B/06A 
command module.
FETCH? does not alter the readings stored in VME memory. Only the *RST 
or INIT
… commands will clear the readings in VME memory. 
The format of readings returned is set using the FORMat[:DATA] command.
Returned Value: REAL,32, REAL,64, and PACK,64, readings are returned in 
the IEEE-488.2-1987 Definite Length Arbitrary Block Data format. This data 
return format is explained in “Arbitrary Block Program and Response Data” on 
page 233. F
or REAL,32, readings are 4 bytes in length. For REAL 64 and PACK, 
64, readings are 8 bytes in length.
PACKed,64 returns the same values as REAL,64 except for Not-a-Number 
(NaN), IEEE +INF, and IEEE -INF. The NaN, IEEE +INF, and IEEE -INF values 
returned by PACKed,64 are in a form compatible with Workstation BASIC and 
BASIC/UX. Refer to the FORMat command for the actual values for NaN, +INF 
and -INF.
ASCii is the default format.
ASCII readings are returned in the form 
±1.234567E±123. For example, 
13.325 volts would be +1.3325000E+001. Each reading is followed by a 
comma (,). A line feed (LF) and End-Or-Identify (EOI) follow the last reading.
Related Commands: MEMory Subsystem, FORMat[:DATA]
*RST Condition: MEMORY:VME:ADDRESS 240000; 
MEMORY:VME:STATE OFF; MEMORY:VME:SIZE 0