298 VT1422A Command Reference  
INPut
When Accepted: Not while INITiated
Related Commands: INP:FILT:FREQ?, INP:FILT:STAT ON | OFF
*RST Condition: generally set to MIN. The VT1529A/B is set to 10 Hz.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
INP:FILT:FREQ  100,(@100:119)
Set cutoff frequency of 100 Hz for first 
20 channels
INPUT:FILTER:FREQ  2,(@15622)
Set cutoff frequency of 2 Hz for RSC 
channel 15622
INPut:FILTer[:LPASs]:FREQuency?
INPut:FILTer[:LPASs]:FREQuency?  (@<channel>) 
returns the cutoff frequency 
currently set for channel. Non-programmable SCP channels may be queried to 
determine their fixed cutoff frequency. If the channel is not on an input SCP, 
the query will return zero.
Parameters
Comments
The <channel> parameter must specify a single channel only.
This command is for programmable filter SCPs only.
Returned Value: Numeric value of Hz as set by the INP:FILT:FREQ command. 
The data type is float32.
When Accepted: Not while INITiated
Related Commands: INP:FILT:LPAS:FREQ, INP:FILT:STATE
*RST Condition: generally set to MIN. The VT1529A/B is set to 10 Hz.
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
Usage
INPUT:FILTER:LPASS:FREQUENCY?  (@155)
Check cutoff freq on channel 55
INP:FILT:FREQ?  (@10024)
Check cutoff freq on RSC channel 0024
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
channel list (string)
100 - 15731
none