VT1422A Command Reference   333
[SENSe]
[SENSe:]DATA:CVTable:RESet
[SENSe:]DATA:CVTable:RESet 
sets all 64 Current Value Table entries to the 
IEEE-754 "Not-a-number."
Comments
The value of NaN is +9.910000E+037 (ASCII).
Executing DATA:CVT:RES while the module is INITiated will generate an error 
3000, "Illegal while initiated."
When Accepted: Not while INITiated
Related Commands: SENSE:DATA:CVT?
*RST Condition: SENSE:DATA:CVT:RESET
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SENSE:DATA:CVT:RESET
Clear the Current Value Table
[SENSe:]DATA:FIFO[:ALL]?
[SENSe:]DATA:FIFO[:ALL]? 
returns all values remaining in the FIFO buffer until 
all measurements are complete or until the number of values returned exceeds FIFO 
buffer size (65,024).
Comments
DATA:FIFO? may be used to acquire all values (even while they are being made) 
into a single large buffer or can be used after one or more DATA:FIFO:HALF? 
commands to return the remaining values from the FIFO.
The format of values returned is set using the FORMat[:DATA] command.
Returned Value: ASCII values are returned in the form 
±1.234567E±123. 
For example 13.325 volts would be +1.3325000E+001. Each value is followed 
by a comma (,). A line feed (LF) and End-Or-Identify (EOI) follow the last value. 
The data type is a string array.
REAL 32, REAL 64 and PACK 64, values are returned in the IEEE-488.2-1987 
Indefinite Length Arbitrary Block Data format. This data return format is 
explained in “Arbitrary Block Program and Response Data” on page 233. 
For REAL 32, each value is 4 bytes in length (the data type is a float32 array). 
For REAL 64 and PACK 64, each value is 8 bytes in length (the data type is a 
float64 array).
Note
Values which are a positive overvoltage return IEEE +INF and a negative overvoltage 
return IEEE -INF (see Table 7-1 on page 294 for actual values for each data format).