VT1422A Command Reference   339
[SENSe]
[SENSe:]FREQuency:APERture?
[SENSe:]FREQuency:APERture? <channel>  
returns the frequency counting 
gate time.
Parameters
Comments
If the channel specified is on an SCP that doesn't support this function, an error 
will be generated. See the SCP's User's Manual for its capabilities.
Related Commands: SENSe:FREQuency:APERture
Returned Value: Returns numeric gate time in seconds. The type is float32.
Send with VXIplug&play Function: hpe1422_cmdReal64_Q(...)
[SENSe:]FUNCtion:CONDition
[SENSe:]FUNCtion:CONDition <ch_list
sets the SENSe function to input the 
digital state for channels in <ch_list>. Also configures digital SCP channels as inputs 
(this is the *RST condition for all digital I/O channels).
Parameters
Comments
The VT1533A SCP senses 8 digital bits on each channel specified by this 
command. The VT1534A SCP senses 1 digital bit on each channel specified by 
this command.
If the channels specified are not on a digital SCP, an error will be generated.
Use the INPut:POLarity command to set input logical sense.
Related Commands: INPut:POLarity
*RST Condition: SENS:FUNC:COND and INP:POL NORM for all digital 
SCP channels.
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
To set second 8-bits of VT1533A at SCP position 4 and upper 4-bits of VT1534A at 
SCP position 5 to digital inputs send:
SENS:FUNC:COND (@133,144:147)
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
channel
string
100 - 163
none
Parameter
Name
Parameter
Type
Range of
Values
Default 
Units
ch_list
string
100 - 163
none