342 VT1422A Command Reference  
[SENSe]
[SENSe:]FUNCtion:CUSTom:REFerence
[SENSe:]FUNCtion:CUSTom:REFerence [<range>,](@<ch_list>) 
links channels 
with the custom Engineering Unit Conversion table loaded with the 
DIAG:CUST:PIECE command. Measurements from a channel linked with 
SENS:FUNC:CUST:REF will result in a temperature that is sent to the Reference 
Temperature Register. This command is used to measure the temperature of an 
isothermal reference panel using custom characterized RTDs or thermistors.
Parameters
Comments
The VT1422A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc, 4 V dc, and 
16 V dc. To select a range, simply specify the range value (for example, 4 selects 
the 4 V dc range). If a value is specified that is larger than one of the first four 
ranges, the VT1422A selects the next higher range (for example, 4.1 selects the 
16 V dc range). Specifying a value larger than 16 generates an error. Specifying 
0 selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. 
The default range (no range parameter specified) is auto range.
If using amplifier SCPs, set them first and keep their settings in mind when 
specifying a range setting. For instance, if the expected signal voltage is to be 
approximately 0.1 V dc and the amplifier SCP for that channel has a gain of 8, 
<range> must be set no lower than 1 V dc or an input out-of-range condition 
will exist.
The *CAL? command calibrates temperature channels based on Sense Amplifier 
SCP setup at the time of execution. If SCP settings are changed, those channels 
are no longer calibrated. *CAL? must be executed again.
Related Commands: DIAG:CUST:PIEC, SENS:FUNC:TEMP, 
SENS:FUNC:CUST:TC, *CAL?
*RST Condition: all custom EU tables erased
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
program must put table constants into array table_block
DIAG:CUST:PIEC 1,table_block,(@108)
send characterized reference transducer 
table for use by channel 8
SENS:FUNC:CUST:REF .25,(@108)
link custom ref temp EU with ch 8
include this channel in a scan list with thermocouple channels (REF channel first)
INITiate then TRIGger module