VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
378 VT1422A Command Reference  
SOURce
To reduce the offset voltage at each dynamic strain "Buffered Output" channel:
1. Measure a Buffered Output channel with its bridge unstrained and place 
the value in a variable arbitrarily called <offset_v>.
2. Send minus <offset_v> to that channel with the SOUR:VOLT command. 
For example: SOUR:VOLT <-offset_v>,(@10000)
Send with VXIplug&play Function: hpe1422_cmd(...)
Usage
SOUR:VOLT .041,(@10031)
correct a -41 mV offset at channel 31 of 
VT1529A/B 00.