VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
402 VT1422A Command Reference  
IEEE-488.2 Common Command Reference
*IDN?
*IDN? 
Identity. Returns the device identity. The response consists of the following 
four fields (fields are separated by commas):
Manufacturer
Model Number
Serial Number (returns 0 if not available)
Driver Revision (returns 0 if not available)
*IDN? returns the following response strings depending on model and options:
HEWLETT-PACKARD,E1422A,<serial number>,<revision number>
The data type for this returned value is string.
Note
The revision will vary with the revision of the driver software installed in the system. 
This is the only indication of which version of the driver is installed.
*LMC?
*LMC? 
Learn Macros query. Returns a quoted string name for each currently defined 
macro. If more than one macro is defined, the strings are separated by commas (,). If 
no macro is defined, *LMC? returns a null string.
*OPC
*OPC 
Operation Complete. Causes an instrument to set bit 0 (Operation Complete 
Message) in the Standard Event Status Register when all pending operations invoked 
by SCPI commands have been completed. By enabling this bit to be reflected in the 
Status Byte Register (*ESE 1 command), synchronization between the instrument 
and an external computer or between multiple instruments can be ensured.
Note
Do not use *OPC to determine when the CAL:SETUP or CAL:TARE commands 
have completed. Instead, use their query forms CAL:SETUP? or CAL:TARE?.
Send with VXIplug&play Function: hpe1422_cmd(...)