54 Field Wiring  
Adding Components to the Terminal Module
The back of the terminal module PCB (printed circuit board) provides surface mount 
pads which can be used to add serial and parallel components to any channel's signal 
path. Figure 2-17 shows additional component locator information (see the 
schematic and pad layout information on the back of the terminal module PCB). 
Figure 2-18 shows some usage example schematics.
Upper layout also applies here
Upper layout also applies here
Lower layout also applies here
Lower layout also applies here
SCHEMATIC OF MODIFIABLE PARTS
HI
HI
LO
LO
TO VT1413C/VT1415A
SH
SL
COMP
LOCATOR
COMP
LOCATOR
Figure 2-17. Additional Component Location 
TO VT1422A
LO
HI
TO USER WIRING
PH
L
PH
PL
0 ohms
0 ohms
LO
HI
Default Circuit
SH
SL
PHL
LO
HI
TO VT1422A
SH
LO
HI
PH
PL
SL
10 kohms
10 kohms
Normal Mode Low-Pass Filter Circuit
TO USER WIRING
LO
HI
SH
0 ohms
SL
0 ohms
PH
L
250
ohm
s
or
200
ohm
s
TO VT1422A
HI
LO
PH
PL
4-20 mA NOTE: input must not exceed common mode limits (usually
±16 volts unless attenuated with a VT1513A SCP)
4 to 20 mA Sense
5 V full scale with 250 ohm (must use 16 volt range)
4 V full scale with 200 ohm (can use 4 volt range for better resolution)
0.
1
F
µ
Figure 2-18. Series & Parallel Component Examples