VXi VT1529A/B 사용자 설명서

다운로드
페이지 529
Programming the VT1422A & VT1529A/B for Remote Strain Measurement   75
Dynamic Strain Port
Offset Control
Each buffered dynamic strain channel includes an offset adjusting DAC 
controlled by the command 
SOURce:VOLTage[:AMPLitude] <-offset_v>,(@<ch_list>)
Reducing the unstrained bridge offset voltage at the dynamic strain port 
channel can allow the VT1432A to measure the channel using a more 
sensitive range. See Figure 3-14 for the offset DAC arrangement.
To reduce the offset voltage at each dynamic strain "Buffered Output" 
channel:
1. Measure an unstrained Buffered Output channel with a 
VT1432A/33B and place the value in a variable arbitrarily called 
<offset_v>.
2. Send minus <offset_v> to that channel with the SOUR:VOLT 
command. For example: SOUR:VOLT <-offset_v>,(@10000)
Note
With a 13 mV resolution, the offset DAC can reduce the Buffered Output 
channel offset to within a few millivolts of zero.
Figure 3-14. Dynamic Strain Offset DAC
+
-
Buffer
To VT1422A
(VT1539A SCP)
X2
Channel Buffer
RJ-45
X2
2:1
Mux
Cal-
Cal+
Optional
Filter
2,10,100Hz
Instrument.
Amplifier
Dynamic Strain
Buffered Output
32:1
Mux
CH31
CH01
CH00
X16
8-bit offset
DAC
SOURce:VOLTage