SECTION 7—TROUBLESHOOTING GUIDE
37
Alarm:
RAPID
Beep....Beep...
Beep...Beep
LOW FLOW ALARM
1a. Kinked or blocked 
tubing, cannula or humid-
ifier.
1b. Flowmeter set at 0.5 
L/min. for Platinum 5 and 
1 L/min. for Platinum 10 
or less.
1a. Inspect for kinks or 
blockages. Correct, clean or 
replace item. Once 
corrected, turn power OFF 
for 60 seconds and then turn 
power back ON.
1b. Reset flowmeter to pre-
scribed flowrate. If prescribed 
flowrate is below 0.5 L/min., 
use the Pediatric Flowmeter 
IRCPF16 (Only with Platinum 
5 and XL).
SYMPTOM:
PROBABLE CAUSE:
SOLUTION: