SECTION 6—PREVENTIVE MAINTENANCE
24
SECTION 6—PREVENTIVE
MAINTENANCE
MAINTENANCE
DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet
before servicing.
before servicing.
NOTE: The Perfecto
2
Concentrators are specifically designed to minimize routine preventive
maintenance at intervals of once per year. In places with high dust or soot levels, maintenance may
need to be performed more often. The following MUST be performed at a minimum of one year in
service to assure years of additional reliability. Refer to Troubleshooting on page 16 for plugged
filter symptoms. Power should be disconnected before beginning preventive maintenance on the
Invacare concentrator.
need to be performed more often. The following MUST be performed at a minimum of one year in
service to assure years of additional reliability. Refer to Troubleshooting on page 16 for plugged
filter symptoms. Power should be disconnected before beginning preventive maintenance on the
Invacare concentrator.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1.
NOTE: There is one cabinet filter located on the back of the cabinet.
1. Remove the filter and clean at least once a week depending on environmental
conditions.
NOTE: Environmental conditions that may require more frequent cleaning of the filters include
but are not limited to: high dust, air pollutants, etc.
but are not limited to: high dust, air pollutants, etc.
2. Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse
thoroughly.
3. Dry the filter thoroughly before reinstallation.
CAUTION
DO NOT operate the concentrator without the filter installed.
FIGURE 6.1 Cleaning the Cabinet Filter