SECTION 6—PREVENTIVE MAINTENANCE
24
SECTION 6—PREVENTIVE 
MAINTENANCE
 DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: The Perfecto
2
 Concentrators are specifically designed to minimize routine preventive 
maintenance at intervals of once per year. In places with high dust or soot levels, maintenance may 
need to be performed more often. The following MUST be performed at a minimum of one year in 
service to assure years of additional reliability. Refer to Troubleshooting on page 16
 for plugged 
filter symptoms. Power should be disconnected before beginning preventive maintenance on the 
Invacare concentrator.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1. 
NOTE: There is one cabinet filter located on the back of the cabinet.
1. Remove the filter and clean at least once a week depending on environmental 
conditions.
NOTE: Environmental conditions that may require more frequent cleaning of the filters include 
but are not limited to: high dust, air pollutants, etc.
2. Clean the cabinet filter with a vacuum cleaner or wash in warm soapy water and rinse 
thoroughly.
3. Dry the filter thoroughly before reinstallation.
CAUTION
DO NOT operate the concentrator without the filter installed.