SECTION 11—REGULATOR
49
FIGURE 11.1   Replacing Regulator
Adjusting Regulator
NOTE: For this procedure, refer to FIGURE 11.1 on page 49.
1. Turn unit on ).
2. Install pressure gauge onto oxygen outlet.
NOTE: Use a 0 ‐ 30 p.s.i. pressure gauge.
NOTE: Check O
2
 pressure at oxygen outlet. It should read a steady five p.s.i. ± 0.5 p.s.i. If pressure 
is not in specification, proceed to STEP 3. If pressure falls within specification, no adjustment is 
needed.
3. Unplug unit.
5. Plug unit in and turn power switch on ( ).
6. Using the access hole located on the control panel, locate the pressure adjustment 
screw in center of pressure regulator top.
7. Insert a 5/32‐inch Allen wrench into the pressure adjustment screw.
8. While reading pressure gauge, do one of the following:
• Turn pressure adjustment screw or adjustment knob, clockwise to increase output 
pressure; or
• Turn pressure adjustment screw or adjustment knob, counterclockwise to decrease 
output pressure.
Location of 
Product Tank
Pressure 
Sensor Tubing
Pressure 
Adjustment Screw 
(5/32-inch Allen 
Wrench)
Pressure Regulator 
(Top)
Pressure 
Regulator
Product 
Tank
Mounting 
Screws
NOTE: For adjusting the regulator there is no 
need to remove the control panel.