SECTION 17—ON/OFF SWITCH
62
SECTION 17—ON/OFF SWITCH
Replacing On/Off Switch
 DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
NOTE: For this procedure, refer to FIGURE 17.1 on page 62.
1. Unplug unit.
4. Remove and code the four spade connectors from back of existing On/Off switch.
5. Compress retaining grips on back of existing On/Off switch and push switch out 
through front of control panel.
NOTE: Ensure proper orientation of the On/Off (
) switch and wiring connections before 
installation. Refer to FIGURE 17.1, Detail “A” and Detail “B” below.
CAUTION
DO NOT install the On/Off (I/O) switch upside down. Universal Off (O) symbol should 
be at bottom and Universal On (I) symbol should be at the top. Possible damage to the 
concentrator may result if not properly installed.
6. Reverse STEP 4 to secure new On/Off (
) switch.
FIGURE 17.1    Replacing On/Off Switch
I/O
I/O
On/Off 
(
) 
 
Switch
I/O
DETAIL “A”
ON
OFF
DETAIL “B”
Spade Connector 
from Brown Wire 
attached here
Brown Jumper 
Wire (From 
Circuit Breaker)
On/Off 
(
) 
 
Switch
I/O
Spade Connectors 
from White Wires 
attached here
Front View
Back View