SECTION 25—SIEVE BED PRESSURE CHECK 
86
SECTION 25— SIEVE BED PRESSURE 
CHECK
Checking Sieve Bed Pressure
 DANGER
To prevent electrical shock, ALWAYS disconnect concentrator from electrical outlet 
before servicing.
Disassembly
NOTE: For this procedure, refer to FIGURE 25.1.
1. Turn power off ( ) and unplug unit. 
3. Turn unit so the back faces you.
4. It may be necessary to remove compressor inlet HEPA filter. Refer to Replacing the 
FIGURE 25.1   Disassembly
Pressure Testing
NOTE: For this procedure, refer to FIGURE 25.2 on page 87.
1. Connect pressure gauge to the ¼ x 2‐inch silicone tubing with connector (p/n 1131392).
2. Remove plug in HomeFill port in back of concentrator.
3. Connect ¼ x 2‐inch silicone tubing with connector to HomeFill port.
/O
Check Valve
Cut PVC Tubing below 
Check Valve to install 
Tap-in Kit
Product 
Tank