Technical Specifications
En
gli
sh
Technical Specifications
Notes: 
The manufacturer reserves the right to change these specifications without notice.
Pressure may be displayed in cm H
2
0 or hPa.
Operating pressure range
4 to 25 cm H
2
O
Maximum single fault steady 
state pressure
40 cm H
2
O
Pressure measurement 
tolerance
±0.5 cm H
2
O ± 4% of the measured reading
Flow measurement tolerance
±0.1 or 20% of reading, whichever is greater
VAuto mode
4 to 25 cm H
2
O (measured at the mask); Min EPAP: 4 cm H
2
O; 
Pressure Support: 0 to 10 cm H
2
O
CPAP mode
4 to 20 cm H
2
O (measured at the mask)
DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871:
Sound pressure level
26 dBA with uncertainty of 2 dBA as measured according to 
ISO 17510-1: 2002
28 dBA with uncertainty of 2 dBA as measured according to 
ISO 17510-1: 2007
Sound power level
36 dBA with uncertainty of 2 dBA as measured according to 
ISO 17510-1: 2007
Dimensions (L x W x H)
 
112 mm x 164 mm x 145 mm
 
Weight
1.3 kg
Power supply
Input range 100–240V, 50–60Hz, 40VA (typical power consumption), 
< 100VA (maximum power consumption)
Housing construction
Flame retardant engineering thermoplastic
Operating temperature
+5
º
C to +35
º
C
Operating humidity
10–95% non-condensing
Storage and transport 
temperature
-20
º
C to +60
º
C
Storage and transport humidity
10–95% non-condensing
Operating altitude
Sea level to 2,600 m
Supplemental Oxygen 
Recommended maximum supplemental oxygen flow: 4 L/min
Electromagnetic compatibility
Product complies with all applicable electromagnetic compatibility 
requirements (EMC) according to IEC60601-1-2, for residential, 
commercial, and light industry environments
Air filter
Two-layered, powder-bonded, polyester non-woven fiber
Air tubing
Flexible plastic, 1 x 2 m
Air outlet
The 22 mm conical air outlet complies with ISO 5356-1
IEC 60601-1 classification
Class II (double insulation), Type CF