AD-4405 Weighing Indicator
 
Specifyi
Specifyi
Specifyi
Specifying 
ng 
ng 
ng the Weighing Range of the First Range
the Weighing Range of the First Range
the Weighing Range of the First Range
the Weighing Range of the First Range    
Step 3  After displaying  Cap for 2 seconds, the first range or the weighing capacity of 
single range will be displayed. When dual range is used, Cap1 is displayed for 2 
seconds. Triangle   1 will be displayed. Specify parameter with the 
0
 - 
9
 keys 
and press the 
ENTER
 key to store it and proceed to the next step. When pressing 
the 
ESC
 key, regardless of what is displayed, the indicator will proceed to the next 
step without changing the parameters. The next step is Zero Calibration in single 
range or the second range resolution in dual range. 
 
 
Specifying 
Specifying 
Specifying 
Specifying the Second Range Resolution
the Second Range Resolution
the Second Range Resolution
the Second Range Resolution    
Step 4  After displaying range2 for 2 seconds, the resolution with decimal point and triangle 
 2 will be displayed. Specify the second range resolution in the same way as the 
first range. The decimal point cannot be moved. Specify the second range resolution 
greater than the first range. Press the 
ENTER
 key to store the parameters and 
proceed to the next step. When pressing the 
ESC
 key, regardless of what is 
displayed, the indicator will proceed to the next step without changing the parameter. 
 
 
Specifying 
Specifying 
Specifying 
Specifying the Second Range Capacity
the Second Range Capacity
the Second Range Capacity
the Second Range Capacity    
Step 5 After displaying Cap2 for 2 seconds, the capacity with unit and decimal point is 
displayed. Specify the capacity in the same way as the first range. The capacity 
should be greater than the first range. Press the 
ENTER
 key to store the parameters 
and proceed to Zero Calibration.   
 
 
5.2.2. 
To Get Stabilized Data 
Step  6  Maintain the following conditions to calibrate the scale (indicator) correctly. 
 
  Maintain a constant temperature, stable power and stable input voltage from the 
load cell. 
 
  Avoid direct sunshine or the near the outlet of an air conditioner.   
 
  Do not install the scale (indicator) where there is a strong magnetic field. 
 
Step  7  Turn the display on and leave it for several minutes.   
 
 
5.2. Calibration Procedure