 key ..... The key to store the parameters temporarily. Proceed to the 
Cal 0
 
display. 
 
Press and hold the 
ON/OFF
 key and press the 
ESC
 key 
 
 
No parameters are changed, CanCel is displayed and the calibration 
mode is finished. 
 
Step18 Press 
the 
ON/OFF
 key to turn the display off. 
 
 
5.3.
5.3.
5.3.
5.3.     Weighing Range Function
Weighing Range Function
Weighing Range Function
Weighing Range Function    
 
  The weighing range function can select "single range" or "dual range". Specify each 
weighing interval (division) for the multi-interval instrument. Each weighing interval is 
displayed according to a net value or gross value. 
Caution 
  When single range is used, performing this function is not required. 
 
 
Example 1 
The gross display. 
 Specified 
parameters: 
 
 
First range 
Range = 50.00kg, division 0.02kg 
 
 
Second range 
Range = 100.00kg (maximum capacity), division 0.1kg 
 Display 
 
 
0kg to 50kg : 
The first range, division 0.02kg. 
 
 
50kg to 100kg : 
The second range, division 0.1kg. 
 
Division 
0.02kg 
          
 Division 
0.1kg 
         0kg                          50kg                            100kg Gross 
      
 
 
 
Weighing 
value 
 
Example 2 
The net display using a 40kg tare value. 
 
Specified parameters: 
The same parameters as example 1. 
 Display 
 
 
-40kg to 50kg : 
The first range, division 0.02kg. 
 
 
50kg to 60kg : 
The second range, division 0.1kg. 
 
Division = 0.02kg 
 
 
 Division=0.1kg 
          -40kg                  0kg                             50kg  60kg  Net 
 
 
 
  Tare value  
                    90kg 100kg  Gross 
 
 
 
40kg                                                          Weighing value 
 
 
 
5.3. Weighing Range Function