Siemens LR 460 사용자 설명서

다운로드
페이지 123
7ML19985JM01
SITRANS LR 460 – INSTRUCTION MANUAL
Page 67
mmmmm
Pa
ra
m
e
te
rs
4.3.9. 20 mA Trim
Calibrates the 20 mA output.
a. Set 
 to MANUAL mode.
b. Set 
 to 20 mA.
c. Attach a calibrated meter and check the output at the terminals. Record the 
remote reading in mA.
d. Enter the recorded value in 
.
e. Restor
 to its previous setting.
f. Confirm that the mA output is as expected.
4.4. Relay Configuration
4.4.1. AIFB
Select AIFB1 or AIFB2.
4.4.2. Function
4.4.3. NC/NO
4.4.4. State
Certificates and Approvals
   Device Certification
Values
Range: 18 to 24 mA 
Related parame-
ters
4.3.1. Function 
 A mA value derived from the ratio of current level to the range 
between Lower Value and Upper Value. 
Handheld 
programmer 
Options
DISABLED
*
MANUAL
LO ALM
HI ALM
ANY ALM
FAIL-SAFE
Handheld programmer 
options
NO *
Normally 
Open
NC
Normally Closed
Handheld programmer 
options
OFF *
Relay 
de-energized
ON Relay 
energized