MODEL 315  HUMIDITY TEMPERATURE METER 
 
 
1.0 Introduction: 
This instrument is a digital Humidity / Temperature meter
 
incorporating a 
polymer capacitive humidity sensor and a semiconductor temperature sensor in 
a compact self-contained enclosure. 
2.0 Specifications: 
Measurement Range: 
Humidity:   
1%RH ~ 99%RH   
Temperature:    -20°C ~ 60°C (-4°F ~ 140°F) 
Accuracy:  at 23°C ± 5°C 
Feature Range Accuracy 
Humidity 
5%RH ~ 95%RH 
±3% RH 
Temperature (°C) 
-20°C ~ 60°C 
±0.8°C 
Temperature (°F) 
-4°F ~ -140°F 
±1.6°F 
Resolution: 
Humidity: 0.1%RH 
Temperature:  0.1°C / 0.1°F 
Response Time:   
Humidity:   
180 sec. in slowly moving air 
Temperature:    10 sec. in slowly moving air 
Display:  
Dual four digit display 
Sample Rate:   
2.5 times per second   
Power:  
9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22 
Battery Life:   
Approx. 85 hours of continuous operation with alkaline 
battery. Low battery indication. 
Operating Condition: 
x Operating Temperature – Humidity: 0°C ~ 40°C (32°F ~ 104°F) @ < 80% RH 
x Storage Temperature – Humidity: -10°C ~ 60°C (14°F ~ 140°F) @ < 70% RH 
x Altitude: Up to 2000 meters (6500 feet). 
Dimension:  
240×54×34 mm (9.5×2.1×1.3 in) 
Weight:  
Approx.180g (6.4 oz) 
Accessories:  
Battery, Instruction Manual 
Options: 
33% RH Humidity standard 
75% RH Humidity standard 
MODEL 315  HUMIDITY TEMPERATURE METER 
 
 
5.4  Notes on Temperature and Humidity Calibrations: 
1.  During either calibration, pressing the        button at any time will return 
the instrument to the normal operating mode, and no calibration data 
will be changed. 
2.  While performing the humidity calibration, the environment should be 
maintained at stable conditions (i.e. constant temperature and constant 
humidity) to optimize the accuracy. 
3.  Always begin the temperature calibration process with the 0°C (32°F) 
temperature-controlled chamber and the humidity calibration process 
with the 32.8% RH@25°C standard cavity. For maximum efficiency, 
insert the probe into the conditioned environment 10 minutes for 
temperature or 30 minutes for humidity before applying power to the 
instrument and starting the calibration process. 
4.  Whenever the probe is inserted into a standard humidity cavity or 
temperature-controlled chamber, always allow at least 30 minutes to 
ensure ambient conditions have stabilized.