Monarch MONARCH 308 사용자 설명서

다운로드
페이지 6
 MODEL 308 MINI DUAL CHANNEL THERMOMETER 
 
 
1.0 Introduction: 
This instrument is a dual channel digital thermometer for use with any Type K 
thermocouple as a temperature sensor. Temperature indication complies with the 
NIST and IEC584 temperature/voltage tables for Type K thermocouples.  
2.0 Specifications: 
Measurement Range: 
-200°C - 1370°C  
-328°F - 2498°F 
Accuracy:  (At ambient of 23 ± 5°C)  
Temperature Coefficient: 
For ambient temperatures from 
0°C - 18°C and 28°C - 50°C, for 
each °C ambient below 18°C or 
above 28°C add the following 
tolerance into the accuracy spec. 
 
0.01% of reading + 0.03°C 
 
(0.01% of reading + 0.06°F) 
 
 
Note: 
The basic accuracy specification does not include the error of the probe.  Please refer to the 
probe accuracy specification for additional details. 
 
Resolution:  
-200°C - 200°C 
 
0.1°C; 800°C 
-1370°C 1°C 
 
-200°F - 200°F  0.1°F; 
outside this range  1°F 
Number of Sensor Inputs:  Two Type K Thermocouples 
Input Protection at Thermocouple Input: 
60V DC or 24Vrms AC 
Display: 
Dual 4 Digit Temperature Displays plus Indicators, see Section 3.0. 
Sample Rate: 
0.6 times per second 
Power Requirement:  9 Volt Battery, NEDA 1604 or JIS 006P or IEC6F22 
Battery Life: 
Approx. 150 hours with alkaline battery. Low battery indication. 
Operating Conditions: 
x Operating Temperature - Humidity: 0°C - 50°C (32°F - 122°F); 0 - 80% RH 
x Storage Temperature - Humidity: -10°C - 60°C (14°F - 140°F); 0 - 80% RH 
x Altitude up to 2000 meters (6500 feet) 
Dimensions: 
164×54×34 mm (6.5×2.1×1.3 in) 
Weight: 
Approx. 180g (6.3 oz) 
Accessories: 
Type K Wire Thermocouple Probe, Battery, Instruction Manual 
 
 
Range Accuracy 
-200°C - 200°C 
±(0.3% reading + 1°C) 
200°C - 400°C 
±(0.5% reading + 1°C) 
400°C-1370°C 
±(0.3% reading + 1°C) 
-328°F - -400°F 
±(0.5% reading + 2°F) 
-200°F - 200°F 
±(0.3% reading + 2°F) 
200°F - 400°F 
±(0.5% reading + 2°F) 
400°F - 2498°F 
±(0.3% reading + 2°F) 