Monarch 306 사용자 설명서

다운로드
페이지 6
 
MODEL 306 
DUAL CHANNEL DATALOGGING THERMOMETER 
 
 
4.0 Operating Instructions: 
4.1  Preparation for Measurement 
Plug the first thermocouple into the socket marked T1, taking care to observe the polarity 
of the thermocouple pins. If dual sensors are being used, plug the second thermocouple 
into the socket marked T2
4.2 Power 
Press the        button to turn the thermometer ON or OFF. 
When first powered on, the LCD will show how much memory 
space is available for use. 
Example:  Display at right indicates 16,000 records of available 
memory. 
4.3  Selecting the Temperature Scale 
When first turned on, the instrument defaults to reading in Celsius (°C). The instrument 
toggles between Celsius (°C) and Fahrenheit (°F) each time the °C/°F button is pressed. 
The instrument remembers the scale setting when last turned off and powers on in that 
setting the next time. 
4.4 Data-Hold 
Function 
The present reading is held on the display by pressing the HOLD button. Pressing the 
HOLD button again releases the hold function and returns the instrument to   
continuous reading.
 
   
In the HOLD mode, the TIMEMAX MIN and °C/°F buttons are disabled, as indicated by 
two consecutive beeps when pressed. 
4.5 Clock 
Setup 
To set the real time clock: 
 
1. Press and hold the MAX MIN button and then power on   
the meter. 
 
2. Press TIME (clock). 
 
3. Press REC (
n) or °C/°F (p) to increase or decrease the number. 
 
4. Press TIME to move to the next item. The adjusting order is: year; 
month; day; hour; minute.     
 
5. Press TIME  again after the last setting to complete the process. 
To abort during a setup process, press the        button. 
4.6 Time 
Function 
Once set up, pressing the TIME button displays time as follows: top of the LCD – year; 
bottom left of the LCD - month and day; bottom right of the LCD - hour and minute.  
Press the TIME button or any other button to exit this mode. This operation will not 
interrupt the recording and MAX/MIN operation. 
4.7  Recording Interval Setup 
To set the recording interval: 
 
1. Press and hold the MAX MIN button and then power on   
the meter. 
 
2. Press HOLD
 
3. Press REC (
n) or °C/°F (p) to increase or decrease the number. 
 
4. Press HOLD to advance to the next item. 
 
5. Press HOLD again after the last setting to complete the process. 
To abort during a setup process, press the        button. 
 
 
MODEL 306 
DUAL CHANNEL DATALOGGING THERMOMETER 
 
 
TX
R X
G N D
4.8 Recording 
Data 
Each momentary press of the REC button will alternately start and 
stop recording. To clear the memory, power off the meter, press 
and hold the REC button followed by the        button, holding both 
on simultaneously for at least 2 seconds. Then release both 
buttons. The display will show "CLR" (as shown to the right) 
indicating that the memory has been cleared. 
4.9 MAX/MIN 
Operation 
Press the MAX MIN button to enter the MAX/MIN mode. In this mode, both the maximum 
and minimum values are simultaneously retained in memory and updated with every new 
data sample. The instrument first enters the MAX mode, and the MAX symbol and value 
are both displayed. Pressing MAX MIN again advances the display to the MIN symbol 
and value. The next press of the MAX MIN button will cause both the MAX and MIN 
symbols to flash. This indicates that the maximum and minimum values have been 
updated in memory and the displayed reading is the present temperature. Each 
successive press of the MAX MIN button circulates the display mode among these 
options. To exit the MAX/MIN mode, press and hold the MAX MIN button for two seconds. 
In the MAX/MIN mode, the °C/°F  button is disabled, as indicated by two consecutive 
beeps if pressed. 
4.10 Auto Power Off 
By default, the instrument powers on in the ‘Auto Power Off’ mode and will automatically 
shut off 30 minutes after the last key operation or RS232 communication.       
To disable this feature, press and hold the HOLD button and then power on the meter. 
Two successive beeps will indicate that ‘Auto Power Off’ is disabled, and the        will not 
be displayed. 
4.11 Low Battery Condition 
When the battery voltage is at or below the minimum for proper operation, the   
 
symbol will show on the LCD indicating that the battery must be replaced. 
4.12 Calibration 
Calibration is conducted by adjusting the following potentiometers to within the   
tolerances shown. 
(Ambient Temperature: 23 ± 3°C) 
Input Adjust 
VR 
Tolerance 
 
0 °C 
VR1 
 
± 0.1 °C 
 
190 °C 
VR2 
 
± 0.1 °C 
  1000 °C 
VR3 
 
± 1 °C 
  1900 °F 
VR4 
 
± 1 °F 
4.13 Digital Output 
The Digital Output is a 9600 bps N 81 serial interface. 
RX is a 5V normally high input port. 
TX is a 5V normally high output port. 
 
Appendix: Thermocouple Probe Specifications   
Model Range  Tolerances 
Description 
TP-K01 
Wire probe 
-50°C to 200°C 
-58°F to 392°F 
±2.2°C or ±0.75% 
±3.6°F or ±0.75% 
Teflon insulation. Maximum insulation 
temperature: 260°C (500°F) 
 
TP-K01: 
Probe for general condition measurements, especially 
for complex and hard to reach places.
 
 
 
 
A simple single point calibration may be 
performed by adjusting VR1 to 0 °C with 
thermally stabilized ice water.