Omega FOB100 사용자 설명서

다운로드
페이지 30
 
10 
 
4  U
SING THE 
T
EMPERATURE 
P
ROBES
 
4.1  Caution 
Each  time  you  connect  a  temperature  probe  to  the  unit,  the  probe  optical connector should be cleaned 
beforehand.
 Otherwise, particles of grease or dirt may obstruct the device internal connector and affect the 
measurements by completely blocking the signal or by  generating too much attenuation when using a long 
fiber length. 
 
Never use a cloth other than the type recommended for fiber optic cleaning. Dampening the cloth with pure 
isopropyl alcohol ensures good cleaning. 
 
Occasionally, clean the inside of the bulkhead connector (attached to the thermometer) with a wipe or cotton 
swab dipped in alcohol (2.5 mm mini foam swabs work best). 
 
0 
 
For very dirty connectors, acetone can also be used, but should be used very carefully, as it is a very strong 
solvent. Care should be exercised so you do not apply any on plastic materials, etc. 
4.2  Description 
The optical connector used is a standard ST type connector. 
 
 
Ferrule
Locking nut
Ferrule key
Device connector
Insert the Ferrule key here 
Pins BNC type
To install the connector,  
Then, turning, insert the two BNC pins into
The appropriate slots on the ST connector.
Slide the ferrule key gently into the above key.