 SITRANS LC 500 – INSTRUCTION MANUAL 
Page 17
mmm
mm
Probe Configur
at
io
n
Electrode Assembly
SITRANS LC 500 electrodes come in a variety of formats to provide the necessary 
characteristics for correct mounting, chemical compatibility, temperature and pressure 
requirements, and dielectric constant of the medium. The main body of the manual 
discusses the standard configuration. Other options, with details, are shown in 
Process Connections
The standard threaded process connection (S-Series) with PFA insulated electrode, 
including the active shield, provides good results in all measurement situations within the 
temperature, pressure, and corrosive capabilities of the materials and seals. This remains 
true over a wide range of dielectric constants in both non-conducting and conducting 
materials.
Any standard process connection is available with the SITRANS LC 500, and special 
versions can be fabricated to match the mounting and application requirements. A wide 
range of threaded and flanged fittings is available. (Contact your local Siemens 
Milltronics representative for details, or check our website at:  
www.siemens-milltronics.com.)
Seal Types
The basic internal seal for the SITRANS LC 500 has a conical-shaped, preloaded 
pressure/leak resistant construction. Up to three levels of seal protection are 
implemented depending on the integrity requirements of the application. A single or 
double cone internal seal forms one or two barriers against leaking, and a third flange 
face gasket is also available in the D and DD seal construction. The flange face seal also 
provides a design with no metal wetted parts if required. 
Process Connection and Seal Configuration of SITRANS LC 500
Process Connection
Seal Type
Seal Description
Threaded
S
Single Cone
Welded Flange
S
Single Cone
Solid Machined Flange
S
Single Cone
D
Single Cone + Teflon flange seal
DD
Double Cone + Teflon flange seal. (Consult your 
local Siemens Milltronics representative.)
SD
Double Cone (used for stilling well applications)