value  = 
Value assigned to key 
 
(DAB)  = 
8 bit data byte in binary format 
 
(^END) = 
Indicates that EOI is asserted with the last byte sent. 
 
 
10.4.4     Establishing Remote Control 
Upon starting execution, the LBA-PC initializes itself in Local Mode. 
The LBA-PC enters Remote Mode when it receives a :REM command or is addressed to listen when 
the REN line is true.  The LBA-PC responds with REM when it receives the :REM command.  There is 
no response when the REN line is used to enter Remote Mode. 
The LBA-PC enters Local Mode when it receives the :LOC command or the REN line is false.  The LBA-
PC responds with LOC when it receives the :LOC command.  There is no response when the REN line 
is used to enter Local Mode.  In Local Mode the LBA-PC receives and responds to remote commands. 
The LBA-PC automatically enters what is called the lockout state when going from Local to Remote 
Mode.  This is equivalent to entering the operator password in the Options | Password menu item.  
While in the lockout state, access is denied to many of the LBA-PC menu items.  The LBA-PC 
automatically exits the lockout state when returning to Local Mode.  You can override the lockout 
state by sending the PSW command or by selecting Options | Password and entering the operator 
password. 
 
Operator’s Manual 
 
LBA-PC 
 
Doc. No. 10654-001, Rev 4.10 
181