Sigma LBA-708 사용자 설명서

다운로드
페이지 281
3.1.7 
 
Generate Gain 
Clicking this item will cause the LBA-PC to execute an automatic Gain Correction calculation cycle.  
The results of this operation will store a gain correction table that will be used to preprocess all data 
frames newly acquired from the Frame Grabber card. 
The status of the Gain Correction condition is visible in the Gain correction Enunciator shown here 
 
and at the bottom of the LBA-PC’s main display screen. 
If the color of the Gain Correction Enunciator is: 
GRAY
  Gain Correction processing is turned OFF. 
GREEN
 The Gain Correction calculation was successful and is OPERATING. 
RED
  Gain Correction processing has been disabled because something caused the usage of it 
to become invalid. 
 
3.1.7.1  How to generate gain 
Before clicking on the Generate Gain... item you must: 
•  Allow the camera to warm up to a normal operating temperature. 
•  Align your laser beam with the camera and set the correct Zoom, Pan Camera type and 
resolution, etc. settings. 
•  Block all light from the detector and perform an Ultracal!. 
•  Uniformly illuminate the camera detector to an intensity level that is about 60% to 90% 
of saturation. 
Now click File, Generate Gain. 
•  After a few seconds.. observe that the Gain Correction enunciator turns Green. 
•  Remove the uniform illumination and re-apply your laser beam to the camera. 
 
Note:  It is highly recommended that you perform a Save Config... and a Save Gain As... operation in 
conjunction with each Generate Gain... operation.  This will insure that you can restore the settings to the 
configuration that will allow the Gain Correction to operate correctly. 
 
3.1.7.2  What is Gain Correction? 
The gain correction algorithm will compute a mean pixel intensity value, and then supply a 
correction factor for each pixel that will correct it to the computed mean.  If your camera has 
dead or seriously defective elements you should not attempt to use or operate with Gain 
Correction. 
 
3.1.7.3  What Disables Gain Correction? 
Gain Correction will become DISABLED if certain data collection conditions, that were in effect 
when the Generate Gain operation was executed, are no longer in effect.  In all cases, these 
conditions are the result of an operators change to the spatial acquisition settings.   The 
DISABLED condition will occur if you make changes to: 