Sigma LBA-708 사용자 설명서

다운로드
페이지 281
If you choose Results Logging, select the Format that you want the data to be logged in. Both 
formats will produce an ASCII text log with comma-delimited entries.  The Spreadsheet format 
will precede the log with a single list of column headings. The Math Program format will 
precede each log entry with a binary number that indicates which results are enabled.  
 
 
 
Figure 9 
 
 
If you choose Export Logging, select the Format that you want the data to be logged in.  The 
.BMP format will produce bitmapped image files.  The .SPA format will produce an ASCII text 
image file with space-delimited entries.  The .CMA format will produce an ASCII text image file 
with comma-delimited entries.  The .CUR format will log only the Cursor image in and ASCII 
format with comma delimited entries.  The .SUM format will log only Column/Row summed data 
in an ASCII format with comma-delimited entries.  See the Export Image section for more 
information regarding these file types. 
 
3.1.11.3 Logging 
Method 
You can choose between three different methods to log data or results.  
•  Continuous:  Logging will commence when you click Start! on the menu tool bar and 
will terminate when you click Stop!, or run out of available disk space. 
•  Frames:  You can limit the logging process to a specified Number of Frames.  Enter 
the number of frames you want to log.  Logging will commence when you click Start! on 
the menu tool bar. 
•  Time:  You can limit the logging process to a specified amount of time.  At the 
HHH:MM:SS entry item, enter the amount of time that you want logging to run.  
Double-click inside the entry item and type in the time value in the format shown.  
Logging will commence when you click Start! on the menu tool bar. 
Operator’s Manual 
 
LBA-PC 
 
Doc. No. 10654-001, Rev 4.10 
41