file record number, and the Number of frames or records to post process. The number 
of frames can be 0, or 1 to the number of frames in the frame buffer or file.  0 means all 
the frames.  The resulting frames will be placed sequentially into the frame buffer, 
beginning at the current frame buffer location. Be careful that you do not overwrite 
something you intended to save.  The post processing toolbar switch 
 is a shortcut to 
selecting the above file name. 
If you are using Post Processing in combination with Logging or Statistics computations, you 
should read Implications of Combining Logging, Statistics. . . in Section 3.1.12. 
 
Operating hints: 
•  When setting up, use the Continuous method with Interval set to 1. 
•  Always Post Process from a File source. 
•  You cannot use Post Process to Ultracal old data.  It must have been calibrated when it was 
captured. 
 
3.2.4.2 Capture 
Interval 
The Interval setting determines the capture rate, up to the fastest rate achievable under the 
present setup conditions.  The value entered in this edit control specifies how many frames will 
occur for each frame captured.  A setting of 1, will attempt to capture every frame from the 
camera; 2, will capture every second frame; 3 every third; and so on... 
Divide this setting into the camera frame rate to compute the rate at which the system will 
attempt to digitize frames, i.e., a 3 will cause a 30 Hz frame rate camera to be digitized at 10 
frames per second.  The ultimate impact of this setting is also determined by the Trigger 
Interval
 setting, see Rate Control in Chapter 5. 
 
Note:  The fastest achievable frame rate is determined by other factors, including the image size, the speed of your 
PC, the type and number of results being computed, the amount of processing being employed. 
 
3.2.4.3 Camera 
The Camera panel in the Capture dialog box permits you to: 
•  Select which camera Input is in use.  Note: This choice is only available if you purchased 
the four-camera option, see section 3.2.4.3.1 below, otherwise the only available choice 
is Camera 1.   
•  Set the Video Gain level of the signal processing system on the frame grabber card. 
Note:  If you have the four-camera option, this control will apply only to the currently 
selected camera.  
•  Set the digitizer Black Level threshold. Note:  If you have the four-camera option, this 
control will apply only to the currently selected camera.   
•  Program the electronic Shutter setting of your camera, if your camera is equipped with 
a remote shutter control feature.  Note:  If you have the four-camera option, this control 
will apply only to the currently selected camera.   
 
Operator’s Manual 
 
LBA-PC 
 
60