Sigma LBA-708 사용자 설명서

다운로드
페이지 281
3.2.5.1 Logging   
This switch will launch you into the Data & Results Logging dialog box. 
 
3.2.5.2 Print   
This switch will cause the selected print options to be printed on your configured printer.  You 
will not get a chance to reconfigure your printer or print setup item when you click this tool, so 
be sure that you have set your options up in File, Print Setup... before you click here. 
 
3.2.5.3 Write 
Protect   
This tool can be used to write protect any frame in the frame buffer.  Before clicking on this tool, 
be sure that the frame number that you want to write protect is indicated in the frame counter.  
The frame counter 
 is located in the lower right corner of the main window.  
When a frame has been write protected, its location in the frame buffer can not be overwritten.  
When new data is flowing into the frame buffer, write protected locations will be skipped. 
To un-write-protect a location, bring the frame number up again and note that the write protect 
button is still in the depressed position; click the tool a second time.  This tool’s button will show 
in the un-depressed state to indicate that the frame is no longer write-protected.  Therefore, the 
only way to determine which frames are write-protected is to cycle through the frames and 
simultaneously view the up/down state of the write-protect button. 
 
Caution:  If you change the frame buffer resolution, all write protected frames and their contents will be deleted. 
 
3.2.6 
Computation...  Energy calibration select results items 
The Computations dialog box is where you select the types of numerical results that you want to 
have calculated and displayed.  You can also calibrate the beam energy here. 
 
Note:  The term energy, shown in
 Italic
, will be used to represent both Energy and Power. 
 
3.2.6.1  Energy of Beam 
The Energy of Beam edit control item lets you calibrate the LBA to the 
energy
 of your laser.  
You must measure the 
energy
 of your beam using an external measuring device, then enter the 
energy
 value here.  Next select the appropriate units in the 
Energy
 Units item.  The value 
entered must be the total 
energy
 of the frame currently displayed.  For accurate results, the 
beam must be completely contained inside the current Hardware Zoom/Pan window, and an 
Ultracal! operation should be performed prior to collecting the frame used for 
energy
 
calibration. 
 
Operator’s Manual 
 
LBA-PC 
 
Doc. No. 10654-001, Rev 4.10 
67