B&K 8510 사용자 설명서

다운로드
페이지 57
 
30
For example: When load receiving one trigger signal, it will switch to 10A current value, and taking 10mS to return the 
current value of 5A.   
10A
5A
Pulsed Transient Operation
10ms
TWD
10ms
TWD
TRIG
TRIG
 
Action  
1. Select the operation mode (
CC,CV,CR or CP), and set a proper value, press 
On/Off
  to turn off the load input. 
2. Press
Shift
+
S-Tran
, set LEVER A=5A, LEVER B=10A, WIDTH B=10mS. Transition mode now is 
CONTINOUS
3. Press ▼ till 
>PULSE
4. Press
Enter
, set transient mode is 
PULSE.  
5. Press
Shift
 + 
Tran
to activate the transient mode. 
6. Press
Shift
  +
Trigger
to start another pulse. Press
Shift
  +
Trigger
, get more pulse.   
7. Press
Shift
 + 
Tran
again to stop the transient operation. 
8. Press
On/Off
,
 the input of load is powered on. 
Toggled Transient Operation 
In this mode, after transition operation start, Load could change the input between the main level and the 
transient level when toggled transient operation is in effect. 
 
For example:   
  When Load receives one trigger signal, Load current will switch between 5A and 10A.   
10A
5A
TRG
TRG
Toggled Transient Operation
 
Steps 
1. Press
Shift
+
S-Tran
, Setting LEVER A=5A, LEVER B=10A, transition mode is TOGGLED.   
2. Press 
Shift
 + 
Tran
to activate the transient mode.