Preparing for Operation 
H25-IR 
 
12
 
Instruction 3015-4342 
2.4 System Setup 
When shipped from the factory, the H25-IR has been configured according to its part number as described 
in Appendix “A”. To change any of the instrument’s system parameters, first enter System Setup by 
pressing both the ENT and ESC buttons on either the instrument’s front panel or Smart probe at the 
same time to display the Main menu. Next, use the Up and Down buttons to highlight “SYSTEM”, and 
then press ENT to display the System Setup menu. 
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 
 
 
 
 
 
 
From the System Setup menu, use the Up and Down buttons to highlight the desired function. Press 
ENT to select that function and display its setup screen. Then perform the appropriate system setup 
procedure as described in Sections 2.4.1 thru 2.4.9. 
 
 
2.4.1 Gas Type 
 
 
 
 
The GAS TYPE function displays a list of gases that the instrument is programmed to monitor. Refer to 
Section 7 Specifications for a complete list of these gases.  
 
Use the Up and Down buttons to highlight the desired gas. Press ENT to save the selection.  
 
 
2.4.2 Leak Units 
 
 
 
 
 
The LEAK UNITS function displays a list of leak rate units of measure that the instrument is capable of 
displaying. Leak rates can be displayed in units of ounces/year (Oz/yr), grams/year (g/yr), milliliters/second 
(mL/s), Pascal Meters
3
/second (PaM/s), or parts per million (ppm).  
 
Use the Up and Down buttons to highlight the desired leak units. Press ENT to save the selection. 
 
 
 
 
 
 
 
 
System Setup Menu:  
 
GAS TYPE 
 LEAK 
UNITS 
 RELAYS 
 LOOP 
 BRIGHTNESS 
 CLOCK 
 PCHK 
MODE 
 PROBE 
BUTTON 
 RESET 
DEFAULTS 
LEAK UNITS 
 
Oz/yr  
 
 
g/yr 
  
mL/s 
  
PaM/s 
  
ppm 
GAS TYPE 
<
 R134A
 
  
 
R401A 
  
R402A 
 < 
R402B