Calibration 
 
Instruction 3015-4342 
27
 
3.4.3 Adjusting the CAL Factor 
Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the 
instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and 
Down buttons to highlight “CAL DATA”, and then press ENT to display the Calibration Data menu. 
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 
 
 
 
 
 
From the Calibration Data menu, highlight “CAL FACTOR” and press ENT to display its setup screen. 
Use the Left and Right buttons to move the blinking cursor across the screen to highlight the digit to be 
modified, and then use the Up and Down buttons to modify that digit. Press ENT to save the new CAL 
factor value and return to the Calibration Data menu. 
 
 
 
 
 
 
 
3.5  Restoring Factory Calibration 
NOTE: 
 Factory calibration cannot be restored if the instrument was calibrated using an 
external leak source. 
 
Restore an instrument to its factory calibration value as follows: 
 
Display the Calibration Data menu by first pressing both the ENT and ESC buttons on either the 
instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and 
Down buttons to highlight “CAL DATA”, and then press ENT to display the Calibration Data menu. 
 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 
 
 
 
 
 
From the Calibration Data menu, use the Up and Down buttons to highlight “RESTORE FACTORY 
CAL”, and then press ENT to display the RESTORE FACTORY CAL screen. 
 
 
 
 
 
Press the ENT button to confirm restoring the instrument to its factory calibration values. 
 
CAL FACTOR  
1
.000 
FOR R134A 
K1=310.89    K2=703.7   K3=1460.3 
K4=0.0    K5=0.0 
Calibration Data Menu:  
 
CAL TYPE 
 EXT 
RATE 
 
EXT GAS TYPE 
 
CAL FACTOR 
 RESTORE 
FACTORY 
CAL 
Calibration Data Menu:  
 
CAL TYPE 
 EXT 
RATE 
 
EXT GAS TYPE 
 CAL 
FACTOR 
 
RESTORE FACTORY CAL 
RESTORE FACTORY CAL 
HIT ENTER TO CONFIRM