4.2.10.2  Clearing the Log 
Display the Log menu as described Section 4.2.10. Next, use the Up and Down buttons to highlight 
“CLEAR LOG” and then press the ENT button to display the Clear Log screen. 
 
 
 
 
 
Press the ENT button to clear the log of all stored measurements. 
 
 
 
 
 
 
4.2.10.3  Viewing the Total of all Logged Measurements 
The ability to display the total of all logged measurements (up to 50) is useful in situations where it is 
necessary to test an installation or piece of equipment that has a maximum permissible total leak rate. 
For example, a halogen fire suppression system can have leaks, but the combined leak rate or the sum of 
all the leaks cannot exceed a certain level, which is typically specified as a fraction of the total charge 
per year.  
 
To display the total of all logged measurements, first display the Log menu as described Section 4.2.10. 
Next, use the Up and Down buttons to highlight “LOG TOTAL” and then press the ENT button to display 
the Log Total screen. Note that the display is auto ranging, causing the decimal point to turn off above the 
value 99.9. In the example below, the total of all logged measurements is 78.5 Oz/yr. 
 
 
 
 
 
 
 
4.2.10.4  Setting the Log Limit 
The ability to set a log limit is useful when using the H-25IR to log the leak rate of a system that has a 
maximum permissible total leak rate. A relay can be programmed to energize when the total of all logged 
measurements (up to 50) equals the log limit (refer to Section 2.4.3 Relays). 
 
To set the log limit, first display the Log menu as described in Section 4.2.10. Next, use thUp and Down 
buttons to highlight “LOG LIMIT” and then press the ENT button to display the Log Limit screen. Use the 
Left and Right buttons to move the blinking cursor across the screen to highlight the digit to be modified, 
and then use the Up and Down buttons to modify that digit. Press ENT to save the log limit value and 
return to the Log menu. 
 
 
 
 
 
 
 
 
 
 
 
 
 
CLEAR LOG 
 
HIT ENTER TO CONFIRM 
CLEAR LOG 
 
CLEARING LOG DATA. 
LOG TOTAL 
 
 COMBINED LEAK RATE 
    = 78.5 Oz/yr 
LOG LIMIT 
IS     99
9