Bacharach H25-IR 사용자 설명서
H25-IR
Maintenance
Instruction 3015-4342
39
5.1.3 IR Emitter
The IR Emitter screen provides readouts of the detector’s voltage, current, resistance, and power
consumption. When contacting the Bacharach Service Department, the operator may be asked to provide
consumption. When contacting the Bacharach Service Department, the operator may be asked to provide
this information to help in the troubleshooting process.
To display the IR Emitter screen, first display the Diagnostics menu as described in Section 5.1. Next, use
the Up and Down buttons to highlight “IR EMITTER” and then press ENT.
To display the IR Emitter screen, first display the Diagnostics menu as described in Section 5.1. Next, use
the Up and Down buttons to highlight “IR EMITTER” and then press ENT.
5.1.4 Pressure
The Pressure screen provides readouts of bench pressure, ambient pressure, the difference between the
two (delta), flow rate through the bench in moles per minute, and the tank pressure of the internal leak
two (delta), flow rate through the bench in moles per minute, and the tank pressure of the internal leak
source. When contacting the Bacharach Service Department, the operator may be asked to provide this
information to help in the troubleshooting process.
To display the Pressure screen, first display the Diagnostics menu as described in Section 5.1. Next, use
the Up and Down buttons to highlight “PRESSURE” and then press ENT.
This screen also allows a technician to manually turn the pump ON and OFF, along with manually
controlling the position of the measure/search flow-rate valve.
Pressing the Down button toggles the pump ON and OFF.
Pressing the Left button opens and closes the measure/search flow-rate valve. An asterisk (∗) on the
display indicates that the valve is in its measure position.
5.2 Factory Menu
The Factory menu is used to set up or troubleshoot the instrument. This menu only appears when
activated by factory personnel or service technicians who know the correct activation procedure.
When activated, display the Factory menu by first pressing both the ENT and ESC buttons on either the
activated by factory personnel or service technicians who know the correct activation procedure.
When activated, display the Factory menu by first pressing both the ENT and ESC buttons on either the
instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and
Down buttons to highlight “FACTORY” and then press ENT to display the Factory menu.
Down buttons to highlight “FACTORY” and then press ENT to display the Factory menu.
Main Menu:
SYSTEM
LOG
SETUPS
SETUPS
CAL
DATA
DIAGNOSTICS
FACTORY
FACTORY
IR EMITTER PARAMETERS
VOLTS=6.1740 mA=129.5
OHMS=47.7 mW=800.0
VOLTS=6.1740 mA=129.5
OHMS=47.7 mW=800.0
BENCH
∗ AMB DELT
11.73 14.04
−2.31
FLOW=0.00950Mpm
LEAK PSI=102.1(100.2)
LEAK PSI=102.1(100.2)
Factory Menu:
AUTO GAIN
DIGIPOT
LEAK
LEAK
RATE
LEAK
GAS
LEAK
TEMP
LEAK
TEMPCO
SENSOR
CAL
MODEL
SELECT
PROBE
TYPE
FIRMWARE