Maintenance 
Instruction 3015-4342 
39
 
5.1.3 IR Emitter 
The IR Emitter screen provides readouts of the detector’s voltage, current, resistance, and power 
consumption. When contacting the Bacharach Service Department, the operator may be asked to provide 
this information to help in the troubleshooting process. 
To display the IR Emitter screen, first display the Diagnostics menu as described in Section 5.1. Next, use 
the Up and Down buttons to highlight “IR EMITTER” and then press ENT
 
 
 
 
 
5.1.4 Pressure 
The Pressure screen provides readouts of bench pressure, ambient pressure, the difference between the 
two (delta), flow rate through the bench in moles per minute, and the tank pressure of the internal leak 
source. When contacting the Bacharach Service Department, the operator may be asked to provide this 
information to help in the troubleshooting process.  
To display the Pressure screen, first display the Diagnostics menu as described in Section 5.1. Next, use 
the Up and Down buttons to highlight “PRESSURE” and then press ENT
 
 
 
 
 
This screen also allows a technician to manually turn the pump ON and OFF, along with manually 
controlling the position of the measure/search flow-rate valve.  
Pressing the Down button toggles the pump ON and OFF. 
Pressing the Left button opens and closes the measure/search flow-rate valve. An asterisk (∗) on the 
display indicates that the valve is in its measure position.  
 
5.2 Factory Menu 
The Factory menu is used to set up or troubleshoot the instrument. This menu only appears when 
activated by factory personnel or service technicians who know the correct activation procedure. 
When activated, display the Factory menu by first pressing both the ENT and ESC buttons on either the 
instrument’s front panel or Smart probe at the same time to display the Main menu. Next, use the Up and 
Down buttons to highlight “FACTORY” and then press ENT to display the Factory menu. 
Main Menu:  
 
SYSTEM 
 LOG 
 SETUPS 
 CAL 
DATA 
 DIAGNOSTICS 
 FACTORY
 
 
 
 
 
 
 
 
 
 
IR EMITTER PARAMETERS 
VOLTS=6.1740     mA=129.5 
 OHMS=47.7        mW=800.0 
BENCH  
∗ AMB        DELT 
11.73       14.04      
−2.31 
       FLOW=0.00950Mpm 
LEAK PSI=102.1(100.2)
Factory Menu:  
 
AUTO GAIN 
 DIGIPOT 
 LEAK 
RATE 
 LEAK 
GAS 
 LEAK 
TEMP 
 LEAK 
TEMPCO 
 SENSOR 
CAL 
 MODEL 
SELECT 
 PROBE 
TYPE 
 FIRMWARE