7 Specifications 
 
Gases Detected..........................Detects and measures, depending on instrument’s part number: 
 
• CFCs, HCFCs, HFCs and Halogens 
  R11, R12, R22, R23, R113, R114, R123, R124, R134a, R227, R236fa, 
  R401A, R402A, R402A, R404A, R407A, R407C, R408A, R409A, R410A, 
  R500, R502, R503, R507, R508B, H1211, H1301, H2402, N1230, FA188 
 
 R600a 
 • 
SF
6
 
 • 
CO
 
• R227 
Detection Method......................Infrared 
Measurement Units..................oz/yr, g/yr, mL/s, PaM
3
/s, ppm 
Measurement Range.................0.03 to 5.00 oz/yr 
0.85 to 142 g/yr 
0.08 x 10
–4
 to 100 x 10
–4 
mL/s 
0.08 x 10
–5
 to 100 x 10
–5
 PaM
3
/s 
0 to 1,000 ppm 
Operating Modes.......................Search: detects presence and location of gas leaks 
Measure: calculates and displays leak rate 
Response Time ..........................0.5 to 1 second (dependent on probe-hose length) 
Sensitivity .................................0.03 oz/yr (0.9 g/yr) all gases 
Resolution..................................0.01 oz/yr (0.1 g/yr) 
Accuracy: 
  Leak Rate .............................±1% of the accuracy value marked on tag attached to leak source 
  PPM ......................................±10% of reading (can be improved to ±1% if recalibrated using a known 
concentration of refrigerant gas) 
Temperature Drift ....................±0.8% of reading per °C 
Calibration ................................Internal: calibration performed using the unit’s built-in leak source 
External: calibration performed using a customer supplied leak  
source of a known gas type and leak rate 
Relays ........................................Four SPDT relays (rated 3 A, 240 VAC), each programmable to energize 
under 11 different operating conditions 
Audio Speaker...........................Audible indication of leak level 
AC Power...................................100 to 240 VAC, 50/60 Hz 
Power Consumption..................15 Watts 
Warm-Up Time .........................2 minutes (30 minutes to meet published specifications) 
Operating Temperature ...........32 to 122 °F (0 to 50 °C) 
Storage Temperature................−4 to 122 °F (−20 to 50 °C) 
Ambient Humidity....................5 to 90% RH, non-condensing 
Dimensions................................4 x 10.75 x 15.5 in. 
(101.6 x 273.1 x 393.7 mm) 