Section 3
Page 17
 
8544install.fm
12 V dc
Before performing any maintenance, disconnect the sampler from the 
power source.
Do not attempt to make any connection or otherwise handle the electrical 
components of the sampler when connected to ac line power if the 
immediate area is wet, or if hands or clothing are wet.
If the circuit breaker or fuse in the ac power source is tripped, determine 
the cause before restoring power to the sampler.
Make sure the power circuit is grounded and protected with a Ground 
Fault Interrupter (GFI).
3.9 12 V dc 
This connection is for supplying power to the sampler controller electronics. 
Power is normally supplied to the sampler controller electronics from either a 
battery or a power supply located in the pocket directly behind the sampler 
controller housing.
In the event of a power outage, an optional 12 V dc battery backup is available 
(Cat. No. 1803). This power backup maintains power to the sampler when ac 
power fails. 
3.10 Auxiliary Connection
The auxiliary interface connector is a general purpose input/output port. Each 
signal is described in 
Table 1 12 V dc Connector Pin Assignments
Pin
Signal Description
A
ground
B
12 to 17 V dc unregulated
Table 2 Auxiliary Pin Assignments
Pin
Signal Description
Wire Color Purpose
Rating
A
12 V dc 
White
This pin can be used to power an external device or flow meter. Must 
be used in conjunction with Pin B (ground).
13.8 V dc 
nominal
B
ground
Blue
Connected to dc ground and is isolated from the earth ground found 
in the ac power line.
C
Flow Pulse Input
Yellow
With the sampler is in Flow Proportional mode and connected to an 
external flow meter, a 5 to 12 V dc input pulse lasting at least 25 
milliseconds will cause the sampler to decrement one count. The 
sampler is capable of accumulating from 1 to 9999 input pulses.
The 12 V dc line found on Pin A can be used directly with a simple 
contact closure to Pin C or an external 5 to 12 V dc pulse may be 
applied providing the ground side of the external signal is connected 
to the sampler ground at Pin B. This count is actuated at the 
beginning of the input signal (the leading edge of the pulse). 
5 to 17 V dc 
pulse with 
duration of 
at least 
25 ms.