Hach 900 사용자 설명서

다운로드
페이지 52
Page 21
 
8544operations.fm
Operation
Section 4
Operation
4.1 Keypad and Key Functions
 shows the 18-position keypad and 24-character alphanumeric 
display. 
Button
Function
ON
Energizes Sampler
OFF
De-energizes sampler and halts program in progress
START PROGRAM
Initiates the sampling program and resets all counts to zero (including datalogger).
When the START PROGRAM button is pressed, all recorded data (program entries and 
collected sample times/dates) are erased from memory. Always download recorded data 
before pressing START PROGRAM.
NEW PROGRAM
Allows input of new sample program...programming choices appear on display.
YES/ENTER
Causes a positive response when a question appears on display; accepts new programmed 
value or previously entered value.
NO/PASS
Causes a negative response when a question appears on display; in standby state, the 
software revision number is displayed.
CLEAR ENTRY
Erases program value shown on display; when held for two seconds, permits enable/disable 
of level actuation/auxiliary control.
DISPLAY FEEDBACK
Displays program status and a review of the program.
CHANGE/HALT
Stops sampler program and places in standby state. When started, press this key prior to 
making any program change and before pressing keys 4, 5, 6, 7, 8, 9, and NEW PROGRAM. 
RESUME PROGRAM
Causes the program to continue from the point at which it is halted.
BOTTLE ADVANCE
Feature not available with this model.
PUMP (Manual Mode)
Pumps sample liquid (pump forward) when held. Pump stops when released.
PURGE (Manual Mode)
Purges sample liquid (pump reverse) when held. Pump stops when released. 
STOP PUMP (Manual Mode)
Stops pump when using PUMP LOCK (see *). When held for 3 seconds (with sampler in a 
standby state), permits selection of other languages. 
TAKE SAMPLE
Initiates sample cycle independent of program.
TIME SET
Allows setting real time and date
TIME READ
Causes real time and date to appear on display. When held for 3 seconds, display indicated 
time and date each sample was taken (or missed). 
* (asterisk)
During programming, causes program to back-step. After CHANGE/HALT, allows altering of 
bottle parameters (container volume, etc.). After CHANGE/HALT, provides review of 
parameters (i.e. number of bottles, volume of bottle(s), line length, etc.). After PUMP/PURGE, 
(manual mode) activates PUMP LOCK. Pump runs forward (KEY 5) or reverse (KEY 6) 
without holding key. KEY 7 stops the pump.