18
Micro Motion
®
 Model 2400S Transmitters 
I/O Wiring – Model 2400S Analog Transmitters
Figure 3-10
Discrete output wiring – External power
Figure 3-11
Output voltage vs. load resistance – Internal power

+
Pull-up resistor or 
DC Relay
3–36 VDC
Maximum sink current: 500 mA
+
Note: Refer to Figure 3-12 for recommended 
resistor versus supply voltage.
CAUTION
Excessive current will damage the transmitter.
Do not exceed 36 VDC input. Terminal current must be less than 500 mA.
0
2
4
6
8
10
12
14
16
18
20
22
24
0
20000
40000
60000
80000
100000
High
 le
v
e
l output v
o
ltag
e (V
o
lts)
Load resistance (Ohms)
Open circuit output
voltage = 24 VDC ±3%
OPERATING REGION