SMV 3000 Transmitter User’s Manual 
1/99
11.4
Diagnostic Messages,  
Continued
Diagnostic Messages,
continued
Table 32
Non-Critical Status Diagnostic Message Table, Continued
SMV Status
SCT Status Message SFC Display Message TDC Status Message
Possible Cause
What to Do
6-4
Output Mode PV1 (DP)
STATUS  TAG ID.#
OUTPUT MODE PV1
OUTPUT MODE PV1
Analog transmitter is operating as a
current source for PV1 output.
Exit Output Mode:
SCT – Press “Clear Output Mode”
button on the DP OutCal tab.
SFC – Press [
OUTPUT
] and [
CLR
]
keys.
6-5
Output Mode PV2 (SP)
STATUS  TAG ID.#
OUTPUT MODE PV2
OUTPUT MODE PV2
Analog transmitter is operating as a
current source for PV2 output.
Exit Output Mode:
SCT – Press “Clear Output Mode”
button on the AP OutCal tab.
SFC – Press [
OUTPUT
] and [
CLR
]
keys.
6-6
Output Mode PV3
(Temp)
STATUS  TAG ID.#
OUTPUT MODE PV3
OUTPUT MODE PV3
Analog transmitter is operating as a
current source for PV3 output.
Exit Output Mode:
SCT – Press “Clear Output Mode”
button on the TEMP OutCal
tab.
SFC – Press [
OUTPUT
] and [
CLR
]
keys.
6-7
Output Mode PV4 (Flow)
STATUS  TAG ID.#
OUTPUT MODE PV4
OUTPUT MODE PV4
Analog transmitter is operating as a
current source for PV4 output.
Exit Output Mode:
SCT – Press “Clear Output Mode”
button on the FLOW OutCal
tab.
SFC – Press [
OUTPUT
] and [
CLR
]
keys.
3-7
PV4 Independent
variable out of range
-
STATUS 3-7
For R250 Laminar Flow transmitters
only.  Asserted when a PV is not
within the range of a term in the
laminar Flow equation.
•  Check the value of every PV
against the ranges in the
Laminar Flow equation.
•  Redefine the equation, if
necessary.