7.  Measurement Procedure
Measurement Procedure
Measurement Procedure
Measurement Procedure    
 
7.1.
7.1.
7.1.
7.1.  Standard Mode Operation
Standard Mode Operation
Standard Mode Operation
Standard Mode Operation    
 
 
  The standard mode can obtain the moisture content with settings of ACCURACY 
and drying program (heating pattern, drying temperature). 
 
7.1.1.
7.1.1.
7.1.1.
7.1.1.  ACCURACY
ACCURACY
ACCURACY
ACCURACY    
 
  ACCURACY of measurement can be set either HI, MID. or LO. 
 
  The sample quantity is automatically selected by ACCURACY. 
 
 The termination value of the analyzing mode is automatically selected by 
ACCURACY and minimum scale value of % display. 
 
  The analysis mode is the program to finish the measurement when a change of 
moisture content per one minute becomes smaller than a preset termination value. 
 
  The settings are as follows: Specify an ACCURACY. 
    
ACCURACY 
 Model 
Minimum 
scale
HI 
MID. 
LO 
  0.001 % 
 0.01 %/min
 0.02 %/min   0.05 %/min
  0.01 % 
 0.02 %/min
 0.05 %/min   0.10 %/min
MS-70 
  0.1 % 
 0.10 %/min
 0.20 %/min   0.50 %/min
  0.01 % 
 0.02 %/min
 0.05 %/min   0.10 %/min
MX-50 
  0.1 % 
 0.10 %/min
 0.20 %/min 
  0.05 % 
 0.05 %/min
 0.10 %/min 
 0.1 
MF-50 
 1 
 0.1 
Preset 
Termination 
value 
ML-50 
 1 
 0.10 %/min
 0.20 %/min 
 0.50 %/min
Sample quantity 
10 g 
5 g 
1 g 
Use Precise 
result 
 Quick 
measurement
 
7.1.2.
7.1.2.
7.1.2.
7.1.2.  Operation 
Operation 
Operation 
Operation     
 
  This operation explanation uses the following example of the MX-50: 
 
  Refer to "8.Selection of Measurement Method" for detail. 
 
  
Input 
Parameters 
 
   
Analysis mode ................................... Standard mode 
 
   
Drying temperature ............................ 130 °C 
 
   
ACCURACY.......................................  LO  
 
   
Sample quantity ................................. Approximately 1 g (Automatic selection) 
 
   
Analysis mode ................................... 0.10 %/min (Automatic selection) 
 
 
  Stored Parameters (Factory Settings for the MX-50) 
 
   
Heating pattern .................................. Standard drying (
 
 
 
   
Measurement unit ............................... Moisture content is based   
on a wet sample 
 
   
Minimum scale value of % display........ 0.01 % 
 
   
Minimum scale value of gram display ... 0.001 g 
 
   
Data memory function........................ Not used 