9-100 Command Reference
:STATus :DREGister0?
80B/81/85
Read Device Status Event Register
This query reads out the contents of the Device Event Register. Reading the De-
vice Event Register clears the register. See Figure 6-14.
Returned format:
<dec.data> = the sum (between 0 and 6) of all bits that are true. See table below:
Bit No.
Weight
Condition
2
4
Last measurement below low limit.
1
2
Last measurement above high limit.
:STATus :DREGister0 :ENABle
80B/81/85
8 <bit mask>
Enable Device Status Reporting
This command sets the enable bit of the Device Register 0.
Parameters:
<dec.data> = the sum (between 0 and 6) of all bits that are true. See table below:
Bit No.
Weight
Condition
2
4
Enable monitoring of low limit
1
2
Enable monitoring of high limit
Returned format:
<bit mask>¿