Agilent Technologies A.18.00 사용자 설명서

다운로드
페이지 581
 
3
Safety Summary
The following general safety precautions must be observed during all phases of 
operation of this instrument. Failure to comply with these precautions or with 
specific warnings elsewhere in this manual violates safety standards of design, 
manufacture, and intended use of the instrument. Agilent Technologies Inc. 
assumes no liability for the customer’s failure to comply with these requirements.
GENERAL
This product is a Safety Class 1 instrument (provided with a protective earth 
terminal). The protective features of this product may be impaired if it is used in a 
manner not specified in the operation instructions.
All Light Emitting Diodes (LEDs) used in this product are Class 1 LEDs as per IEC 
60825-1.
This product has been designed and tested in accordance with IEC Publication 
1010
, "Safety Requirements for Electronic Measuring Apparatus," and has been 
supplied in a safe condition. This instruction documentation contains information 
and warnings which must be followed by the user to ensure safe operation and to 
maintain the product in a safe condition. 
ENVIRONMENTAL CONDITIONS
This instrument is intended for indoor use in an installation category II, pollution 
degree 2 environment. It is designed to operate at a maximum relative humidity of 
95% and at altitudes of up to 2000 meters. Refer to the specifications tables for the 
ac mains voltage requirements and ambient operating temperature range.
Ventilation Requirements: When installing the product in a cabinet, the convection 
into and out of the product must not be restricted. The ambient temperature 
(outside the cabinet) must be less than the maximum operating temperature of the 
product by 4° C for every 100 watts dissipated in the cabinet. If the total power 
dissipated in the cabinet is greater than 800 watts, then forced convection must be 
used.
BEFORE APPLYING POWER
Verify that the product is set to match the available line voltage, the correct fuse is 
installed, and all safety precautions are taken. Note the instrument's external 
markings described under Safety Symbols.