Changing the Patient Category
Setting up your Monitor   3-49
S
e
tt
ing up y
our
 
M
onit
o
r  
1
The figure given for overpressure protection is the maximum pressure 
the cuff can reach for 2 seconds before it deflates. 
NBP Recommendations
Philips Medical Systems makes the following recommendations when 
selecting a patient category for NBP:
1. Choose the category with the most appropriate safety values for 
your patient.
2. Reduce the stress on the patient by choosing the appropriate cuff 
size (a cuff that is too large prolongs inflation time).
3. Use a repetition time appropriate to the patient. 
ECG
The following factors can cause the cardiotach to calculate an incorrect 
heart rate: 
 Morphology of the ECG, 
 Weak signal, 
 Paced patients. 
Morphology of
the ECG
Philips offers two different ECG algorithms to allow for differing 
morphologies. 
Max. deflation time
 80s
80s
60s
Max. measurement time
100s
100s 
60s
Overpressure protection
1
 
300 mmHg 
300 mmHg
150 mmHg 
Cuff Size    
A to E 
Any Cuff
1 to 4 & A
Default Inflation Pressure
165 mmHg 
125 mmHg
100 mmHg 
NBP Safety Values for Differing Patient Categories
Safety Value
Patient Category
Adult
Pedi
Neo