Chapter 1
Taking Measurements with the NI 4050
1-8
www.ni.com
2.
Select diode mode.
3.
Select the range for your measurement. Only the 2 V range is available 
for diode measurement.
The value indicator will indicate the voltage drop measured. If the display 
indicates 2.200 VDC, the diode is either reverse biased or defective. See the 
 section of Chapter 2, 
, for more 
information on diode measurements.
Measure Current
You can use the NI 4050 to measure current with an optional National 
Instruments CSM series current shunt module. These accessories are 
connected between the NI 4050 cable and the test probes as shown in 
Figure 1-7.
Figure 1-7.  Connections for Current Measurement
Current shunt accessories contain a precision resistor that converts the 
current through the shunt into a voltage that the NI 4050 can measure in 
voltage mode. While the soft front panel cannot measure current directly 
with the NI 4050, you can calculate the value of the current flowing through 
the shunt by dividing the voltage measured by the value of the precision 
resistor.
Current
Source
Current Shunt
Accessory
250 V
MAX.
LO
HI