3.2  D/A Calibration Steps 
 
1.  Run A8111CAL.EXE 
2.  Connect D/A channel0,, pin 30 of CN1 to DVM   
3.  Adjust VR1 until DVM=5.0000V 
4.  Press “Enter" key   
5.  Adjust VR2 until DVM=0.0000V 
 
 
3.3  A/D Calibration Steps 
 
1. Press “Enter" key 
2. Connot A/D Channel 0 to analog ground, CN1-Pin1 to CN1-Pin20. 
3. Input stable DC 4.9988V to A/d channel 1, pin2 of CN1. 
4. Connect DVM to TP1(-) & (+). 
5. Adjust VR5 until DVM=0.0000V. 
6. Press <Enter> Key. 
7. Adjust VR4 until A/D data shows 4094 or 4095. 
8. Press <Enter> Key 
9. Adjust VR3 until A/D data shows 2047 or 2048. 
10.  Repeat step_3 to step_11 until: 
A/D channel 0 input 4.9988VÎA/D reading 4094 or 4095. 
A/D channel 0 input 0 V      ÎA/D reading 2047 or 2048. 
11.  Press <Enter> Key. 
 
 
 
3.4   DI/O Testing 
12.  Use 20-pin flat cable (OME-CA-2010, OME-CA-2020 Option) to link CN2 to CN3. 
13.  Press <Enter> Key. 