Start manual data recording:
Shortcut: Press and hold the yellow ESC key and press the 
 
key once for each desired 
measurement. 
– Or, use the StorE menu (with assignment of a name to the memory block):
Key Sequence: SEt 
 
StorE 
 LAbEL 
 
no 
 
YES 
 
mybloc 
 
 
ESC 
Settings:
mybloc = name of the current memory block (optional) 
Start recording with the ESC key.
A new memory block is activated for the storage of data. The new memory block can be 
named if desired (LAbEL). The name is made available for purposes of analysis with 
METRAwin 10 software. After recording has been activated with the ESC key, the following 
display appear: 
bL.nnn = ID number for the current memory block between 1 and 128 (continue with  )
d.nnnn = ID number for the current measurement value within the memory block between 1 
and 1000 (continue with  )
nn.n = percentage memory occupancy from 00.1 to 99.9%
The display can be returned to the initial menu with the ESC key without causing interference 
to memory mode operation.
Interrupt Data Recording
Key Sequence: Activate the ESC key twice.
Measurement values can now be queried with the reCAII and SEE functions. 
Data recording can be resumed by pressing and holding the ESC key, and simultaneously 
pressing the 
 key.
Stop Data Recording (StorE mode)
Key Sequence: SEt 
 
StorE 
 LAbEL 
 
no 
 
bL.nnn 
 
StOP 
Settings:
StOP = end of data recording
Manual data recording is stopped and the corresponding memory block is closed off.
Shortcut: Simultaneously press the yellow ESC key and the 
 key.
7.9
View Data at the Adapter – reCAll
Statistical Overview of Stored Data
Key Sequence: rECAll  
bbbb    
StAt 
bEGin 
 
dd  
hh:mm  
mm:ss  
End  
dd  
hh:mm  
mm:ss  
MIN xxxxx 
 
dd
 
 
hh:mm  
mm:ss  
MAX xxxxx 
 
dd
 
 
hh:mm  
mm:ss 
Settings:
bbbb = block number for stored data
bEGin = time recording was begun for data in the memory block
End = time recording was ended for data in the memory block
dd = weekday
hh:mm, mm:ss = time in hours minutes and seconds
MIN, MAX = display minimum and maximum values
xxxxx = measurement value (digital and analog display)
This function displays a statistical overview for a selectable memory block. Memory block ID 
number, beginning and ending time of the recording, minimum and maximum values, as well 
as the date and time for each, appear at the display.